Inspiratory pressure (Pressure) in the One-
step recruitment maneuver
Setting range
Increment value
Accuracy
Highest inspiratory pressure (Pinsp max) in
the Multi-step recruitment maneuver
Setting range
Increment value
Accuracy
Highest PEEP (PEEP max) in the Multi-step
recruitment maneuver
Setting range
Increment value
Accuracy
Pressure amplitude above PEEP: ∆Pressure
in the Multi-step recruitment maneuver
Setting range
Increment value
Accuracy
Minimum pressure at patient port in accor-
dance with ISO 80601-2-13
15.8
Breathing system
Total volume (when applying a maximum
tidal volume of 1500 mL)
Without CO
absorber
2
With CLIC absorber 800+ disposable
CO
absorber
2
With reusable CO
absorber and
2
Drägersorb 800+
Information
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
PEEP +1 to 80 hPa (cmH
(3 to 80 hPa (cmH
O) when PEEP = Off)
2
1 hPa (cmH
O)
2
±10 % of the set value or ±2 hPa (cmH
value applies)
15 to 80 hPa (cmH
O)
2
1 hPa (cmH
O)
2
±10 % of the set value or ±2 hPa (cmH
value applies)
PEEP to 35 hPa (cmH
2
(2 to 35 hPa (cmH
O) when PEEP = Off)
2
1 hPa (cmH
O)
2
±10 % of the set value or ±2 hPa (cmH
value applies)
5 to 30 hPa (cmH
O)
2
1 hPa (cmH
O)
2
±10 % of the set value or ±2 hPa (cmH
value applies)
–3 hPa (cmH
O)
2
2.18 ± 0.2 L
3.53 ±0.2 L
3.67 ±0.2 L
The total volume of the breathing system is typically
smaller and is calculated based on the set tidal volume
plus a reserve volume for compliance compensation.
Technical data
O)
2
O) (the greater
2
O) (the greater
2
O)
O) (the greater
2
O) (the greater
2
279