Advertisement

Operation
8.2.3
Setting and starting the therapy
1. Select the ventilation mode (1).
The following ventilation modes are available:
– Man/Spon
– VC - AF
– VC
– PC
– PSV
The following operation modes are also available:
– Ext. FGO
– Pause
– Monitoring
For further information see: "Description of the ventilation modes", page 308.
WARNING
Risk of patient injury
The use of minimal-flow settings or low-flow settings can lead to the following:
– Accumulation of metabolic by-products in the breathing system
– Condensed water in hoses
– Condensed water in the piston diaphragm
► Check the hoses for condensation at regular intervals. Remove the condensed
water if required.
► When repositioning the patient or changing the hose arrangement, check the
hoses for condensation. Remove the condensed water if required.
► Before a change of patient, always perform the following measures:
– Remove the condensed water.
– Perform a leakage test or a complete system test.
► Prevent condensed water from entering the patient's airways. To do so, take at
least one of the following measures:
– Equip the Y-piece with an HME or HMEF.
– Use water traps in the breathing hoses. Position the water traps at the lowest
– Position the patient's head higher than the ports on the breathing system.
► Follow the recommendations of professional societies.
1) Pause must be activated in the system setup.
2) Only for devices with integrated PGM. Pause must be deactivated in the system setup.
146
Select mode
5
4
3
1
1)
2)
point of the breathing hoses.
Instructions for use
2
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n

Advertisement