Dräger Atlan A100 Instructions For Use Manual page 248

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Troubleshooting
Priority
Alarm
!!! 220
Apnea
!! 255
Apnea
!!! 220
Apnea (no CO
!! 255
Apnea (no CO
!!! 220
Apnea (no flow)
!! 255
Apnea (no flow)
248
Cause
No breathing or ventilation.
No breathing or ventilation.
)
No breathing or ventilation.
2
Sample line is not con-
nected.
)
No breathing or ventilation.
2
Sample line is not con-
nected.
No breathing or ventilation.
Fresh-gas flow is insuffi-
cient, breathing bag is
blocked or positioned incor-
rectly.
The breathing hoses are
blocked or leaking.
No breathing or ventilation.
Fresh-gas flow is insuffi-
cient, breathing bag is
blocked or positioned incor-
rectly.
The breathing hoses are
blocked or leaking.
Instructions for use
Remedy
Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
Connect sample line to
breathing circuit.
Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
Connect sample line to
breathing circuit.
Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
Check fresh-gas settings and
position of breathing bag.
Check breathing circuit and
tube.
Start manual ventilation!
Check ventilation settings.
Check spontaneous breath-
ing ability of the patient.
Check fresh-gas settings and
position of breathing bag.
Check breathing circuit and
tube.
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n

Advertisement

Table of Contents
loading

Table of Contents