5.3.2
Canceling the setting procedure or the change procedure
If a change to a parameter should be canceled (color is still yellow), the following
options exist to retain the previous setting:
● Touch the changed parameter again. This resets the selection of and the change
to the parameter.
● Select another parameter. This selection resets the change made to the
previous parameter.
● Do not press the rotary knob. After 15 seconds, the change is reset and signal
tones sound during the last 5 seconds (timeout).
5.3.3
Activation of buttons
Some buttons are immediately active without additional confirmation. The color
immediately turns to dark green.
Examples:
– Selecting a view
– Deactivating the CO
5.3.4
Operating the flow control valves
The flow control valves of the mechanically controlled gas mixers and the O
flowmeter are operated as follows:
Opening the flow control valve
● Turn the flow control valve counterclockwise.
Closing the flow control valve
● Turn the flow control valve clockwise to the end stop.
In the subsequent chapters of this document, the following is represented by
simplified explanations:
– "Open the flow control valve."
– "Close the flow control valve."
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
alarms