Dräger Atlan A100 Instructions For Use Manual page 206

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Configuration
11.5.3
System setup > Alarms
11.5.3.1
Vertical tab "Limits"
Headline/
Parameter
19 to 99;
FiO
[%]
2
18 to 98
FiO
[%]
2
etCO
2
0.1 to 9.8;
[%]; [kPa]
1 to 75;
[mmHg]
etCO
2
[%]; [kPa]
0.0 to 9.7;
[mmHg]
0 to 74
inCO
2
0.1 to 1.4;
[%]; [kPa]
1 to 10;
[mmHg]
5 to 99
Paw
[mbar]; [hPa];
[cmH
O]
2
0.10 to 9.95
inSev
[%]; [kPa]
inSev
0.00 to 9.85
[%]; [kPa]
0.1 to 20.0
inDes
[%]; [kPa]
inDes
0.0 to 19.9
[%]; [kPa]
0.10 to 9.95
inEnf
[%]; [kPa]
inEnf
0.00 to 9.85
[%]; [kPa]
206
Setting range
19 to 99;
Off
Off
18 to 98
20
20
0.1 to 9.8;
Off
Off
7.0
7.0
1 to 75;
Off
Off
53
53
Off;
Off;
0.0 to 9.7;
Off;
Off;
0 to 74
0.1 to 1.4;
Off
Off
1.1
1.1
1 to 10;
Off
Off
8
8
5 to 99
40
25
0.10 to 9.95
4.40
5.10
Off;
Off;
0.00 to 9.85
0.1 to 20.0
12.5
14.5
Off;
Off;
0.0 to 19.9
0.10 to 9.95
3.60
4.10
Off;
Off;
0.00 to 9.85
Instructions for use
Description
19 to 99;
Inspiratory oxygen fraction
Off
90
18 to 98
20
Expiratory CO
0.1 to 9.8;
Off
7.0
1 to 75;
Off
53
Off;
0.0 to 9.7;
Off;
0 to 74
Inspiratory CO
0.1 to 1.4;
Off
1.1
1 to 10;
Off
8
5 to 99
Airway pressure
20
0.10 to 9.95
Sevoflurane
6.70
Off;
0.00 to 9.85
0.1 to 20.0
Desflurane
19.0
Off;
0.0 to 19.9
0.10 to 9.95
Enflurane
5.40
Off;
0.00 to 9.85
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
concentration
2
concentration
2