Dräger Atlan A100 Instructions For Use Manual page 179

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Manual setting
In the Alarms dialog, the alarm limits can be set either manually or automatically.
1. Open the Alarms dialog.
2. Open the Limits tab (1).
3. Set the upper alarm limit (2).
4. Set the lower alarm limit (3).
Automatic setting
Alarm limits can be automatically adapted to current measured values or set values.
1. Open the Alarms > Limits dialog.
2. To adjust the alarm limits for one individual parameter, touch and confirm one of
the Autoset buttons (5).
To adjust the alarm limits for all parameters, touch and confirm the Autoset all
button (4).
The function for adjusting all alarm limits can be called up directly with the Autoset
limits button in the main menu bar.
Only use the automatic adjustment when measured values and set values are
stable.
The lower alarm limit for the xMAC level is also adjusted during automatic setting,
see page 180.
For the configuration and algorithm, see page 208.
8.10.2
Activating and deactivating CO
Prerequisite: The device is equipped with the "integrated patient-gas measurement
module".
CO
2
deactivated.
1. Open the Alarms dialog.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Alarms
1
5
monitoring (alarms for inCO
4
alarms
2
, etCO
, and CO
apnea) can be activated or
2
2
2
Operation
2
3
179