Dräger Atlan A100 Instructions For Use Manual page 251

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Priority
Alarm
!!! 120
Breathing system failure
!! 255
Breathing system failure
!!! 255
Breathing system temp.
high?
!
100
Breathing system warmer
failure
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Cause
The breathing system is not
correctly installed or has
leakage.
The breathing system is not
correctly installed or has
leakage.
The breathing system
warmer is faulty.
The breathing system
warmer is faulty.
Troubleshooting
Remedy
Check patient condition.
Check the breathing system.
Verify that the sensor port for
the O
sensor is sealed.
2
If the problem persists, use a
manual resuscitator.
When the case has been
completed, perform a sys-
tem test.
Check if the piston dia-
phragm is correctly inserted.
If the problem persists,
replace the breathing system.
Check patient condition.
Check the breathing system.
Verify that the sensor port for
the O
sensor is sealed.
2
If ventilation is impaired, use
a manual resuscitator.
When the case has been
completed, perform a sys-
tem test.
Check if the piston dia-
phragm is correctly inserted.
If the problem persists,
replace the breathing system.
Check the inspiratory breath-
ing gas temperature as close
to the Y-piece as possible.
Remove breathing system
cover.
Use longer inspiratory hose.
Open the system setup and
turn off the breathing system
warmer.
If necessary, proceed as fol-
lows: -Turn off the device with
the main switch, then venti-
late the patient with the
breathing bag. Or -Discon-
nect the patient, then use a
manual resuscitator.
Check breathing circuit for
condensation. Increase
fresh-gas flow if necessary.
Call Dräger.
251