Dräger Atlan A100 Instructions For Use Manual page 64

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Assembly and preparation
The flow sensor can ignite medications or other substances based on highly
flammable substances. As a result, user and patient could be put at risk.
► Do not nebulize medications or other substances that are easily flammable or
spray them into the device.
► Do not use substances containing alcohol.
► Do not allow flammable or explosive substances to enter the breathing system
or the breathing circuit.
► Do not use cyclopropane or ether.
Pressure reducers
Pressure reducers have an internal pressure release valve. If a fault occurs, gas
may escape into the ambient air. Personal injury or property damage may result as
a consequence.
► Do not block or cover the pressure release valve.
4.1.6
Mechanical safety
Accessories
If the weight of the accessories is unevenly distributed on the device or exceeds the
permissible limits, the device may tip over. The end stops of the support arms in use
must be functional, otherwise the arms may swing unchecked.
► Distribute the weight evenly.
► Pay attention to the maximum weight on each support arm.
► Check whether the additional counterweight in the lower section of the trolley is
required (see "Compact version with counterweight", page 68).
► Check the functional integrity of the support arm end stops after the following
activities:
– After fitting accessories
– After transporting the device
Trapping of body parts
Movable components or attached parts may cause crushing due to trapping. Pay
special attention to edges, movable parts, and corners when working with the
following components:
► Breathing system cover
► Drawers
► Folding table extension
► Pull-out writing tray
► Support arms for mounted devices
► Accessories such as gas cylinders, vaporizers, CLIC absorbers, and CLIC
adapters
Accidental movement of the trolley
An unbraked device may accidentally move during operation. As a result, user and
patient could be put at risk.
► To prevent this, actuate the castor brakes. Check the functional integrity of the
brakes.
64
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n