3.2.12
Safety functions
3.2.12.1
Emergency O
CAUTION
Risk of increased anesthetic agent delivery
When the emergency O
be delivered into the breathing system in accordance with the vaporizer setting.
When the emergency O
flow anesthesia, an increased quantity of anesthetic agent may enter the breathing
system. This may lead to an increased anesthetic gas concentration.
► Carefully monitor the gas mixture.
1. Check the vaporizer setting.
2. Set the O
WARNING
Risk of fire
In combination with oxygen or nitrous oxide, ignition sources such as
electrosurgical devices and laser surgical devices can cause fires.
► If ignition sources are present, do not open the flow control valve on the O
flowmeter. Leave the flow control valve completely closed.
3. Open the flow control valve (2) on the O
This O
3.2.12.2
Backup manual mode
In various technical fault situations, the backup manual mode enables a direct
changeover to manual ventilation in order to continue the therapy. When the backup
manual mode is activated, an acoustic and optical alarm signal is issued with high
priority. The alarm is automatically downgraded to low priority after 20 seconds (see
"Alarm delay, alarm escalation, and alarm deescalation", page 198).
50
delivery (electronically controlled gas mixer)
2
delivery (Add. O
2
delivery is used during low-flow anesthesia or minimal-
2
switch (1) upwards to the Add. O
2
1
flow flows through the vaporizer.
2
Instructions for use
) is in use, anesthetic agent continues to
2
position.
2
2
flowmeter and set the desired flow.
2
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
2