Dräger Atlan A100 Instructions For Use Manual page 292

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Technical data
Nominal voltage
Fuse
Current
Backup time with fully charged battery
During mechanical ventilation (VC - CMV,
VT = 500 mL, RR = 10 /min, I:E = 1:2,
PEEP = 5 hPa, FG flow = 10 L/min O
with less than 5 completed battery dis-
charge cycles
During mechanical ventilation (VC - CMV,
VT = 500 mL, RR = 10 /min, I:E = 1:2,
PEEP = 5 hPa, FG flow = 10 L/min O
with less than 20 completed battery dis-
charge cycles
During ventilation in Man / Spon mode
and with less than 20 completed battery
discharge cycles
Charging time (to achieve at least 30 minutes
backup time)
Charging power
Gas supply
Gas quality requirements
Oil content
Dew point
Particle size
Supply pressure for O
2
Maximum short-term peak inlet flow with elec-
tronically controlled gas mixer at 6.9 kPa x 100
[bar] (100 psi or 0.69 MPa) supply pressure
O
2
Air
Without bronchial suction system
Including a directly connected bronchial
suction system
N
O
2
Distribution piece for central O
Maximum permissible flow
Distribution piece for central Air supply
Maximum permissible flow
Drive gas
292
) and
2
) and
2
, Air, N
O
2
supply
2
Instructions for use
24 V
F15A 80V UL248-14, breaking capacity 1000 A, size
19.7 mm x 19 mm x 5 mm
Maximum 15 A
90 min or longer
Typically 120 min
45 min or longer
90 min or longer
At least 8 hours
Maximum 50 W
3
<0.1 mg/m
5 °C (41 °F) at ambient temperature
Dust-free air (filtered with pore size <1 μm)
2.7 to 6.9 kPa x 100 [bar]
39 to 100 psi
0.27 to 0.69 MPa
135 L/min (applies only when there is no distribution
piece for the central O
2
50 L/min
130 L/min
40 L/min
20 L/min
70 L/min
Not needed
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
supply)