Technical data
15.4
Fresh-gas delivery with electronically controlled gas
mixer
Standardization
O
concentration FG O
2
Setting range
Increment value
Accuracy
Fresh-gas flow FG flow
Setting range
Increment value
Accuracy
Information
15.5
Fresh-gas delivery with a mechanically controlled gas
mixer and electronic flow measurement
Standardization
O
concentration
2
Setting range for fresh-gas flow
Information
Electronic measurement of fresh-gas flow
Range
Accuracy
Resolution of the value displayed on the
screen
274
All data are standardized to STPD conditions.
2
21 to 100 Vol% (carrier gas: Air)
25 to 100 Vol% (carrier gas: N
1 Vol%
±5 % of the set value or ±2 Vol% (the greater value
applies)
Off; 0.20 to 15.00 L/min
0.05 L/min (below 1.00 L/min)
0.10 L/min (between 1.00 and 5.00 L/min)
0.50 L/min (above 5.00 L/min)
±10 % of the set value or ±50 mL/min (the greater
value applies)
The maximum flow dispensed can be reduced depend-
ing on the ambient pressure or temperature. At the lim-
its of the permissible operating conditions, a rate of at
least 13.5 L/min is reached. (An STPD flow of
13.5 L/min corresponds to an ATPD flow of 21 L/min at
650 hPa ambient pressure, for instance.)
All data are standardized to STPD conditions.
21 to 100 Vol%
Off to at least 12.0 L/min (O
The specified setting range can be reduced depending
on the ambient pressure or temperature. At the limits of
the permissible operating conditions, a rate of at least
10 L/min is reached. (An STPD flow of 10 L/min corre-
sponds to an ATPD flow of 15.6 L/min at 650 hPa
ambient pressure, for instance.)
0.00 to 15.00 L/min (O
±10 % of the delivered flow or ±0.12 L/min (the greater
value applies)
0.01 L/min (from 0.00 to 0.20 L/min)
0.02 L/min (from 0.20 to 0.50 L/min)
0.05 L/min (from 0.50 to 1.00 L/min)
0.10 L/min (from 1.00 to 15.00 L/min)
Instructions for use
2
, Air, and N
2
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
O)
2
, Air, and N
O)
2
O)
2

Advertisement