Operation
2. Select the anesthetic gas to be used (3):
– Auto: The flow measurements are automatically corrected for the anesthetic
– Des: The measured flow values will be corrected for desflurane using an
– Sev: The measured flow values will be corrected for sevoflurane using an
– Other: Select this option if you want to use an anesthetic gas other than
– None: The measured flow values will not be corrected.
The following step 3 applies only to main devices with mechanically controlled
gas mixer with flow tubes:
3. Select the carrier gas to be used (5):
– Auto: The measured flow values are automatically corrected for the carrier
– Air: The measured flow values will be corrected for Air.
– N
8.3.5
Changing the ventilation mode
1. In the therapy bar, touch the tab (1) of the new ventilation mode.
When the ventilation mode is changed, the ventilation settings are adopted from
the previous ventilation mode or they are sensibly derived. In addition, the alarm
settings are adjusted to reasonable values, see page 196.
2. Adjust the therapy controls (3) or the buttons (2).
3. Activate the ventilation mode with the rotary knob. A signal tone is emitted when
the mode is changed.
150
gas detected by the anesthetic gas monitor. If no anesthetic gas is detected
or if the data connection to the anesthetic gas monitor is interrupted, no flow
correction takes place and the message None (4) is displayed.
average value. For more information, see the table on page 214.
average value. For more information, see the table on page 214.
desflurane or sevoflurane. The measured flow values will not be corrected.
gas detected by the anesthetic gas monitor. If no carrier gas is detected or if
the data connection to the anesthetic gas monitor is interrupted, flow
correction will be performed for Air.
O: The measured flow values will be corrected for nitrous oxide.