Dräger Atlan A100 Instructions For Use Manual page 199

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Alarm
O
measurement not
2
available
N
O measurement not
2
available
Agent measurement
not available
O
measurement tem-
2
porarily inaccurate
CO
sensor accuracy
2
low
Agent measurement
temporarily inaccurate
N
O measurement
2
temporarily inaccurate
Measured gas concen-
trations temporarily
inaccurate
Gas measurement alarms with inspiratory O
Alarm
FiO
low
2
FiO
high
2
O
measurement not
2
available
O
measurement tem-
2
porarily inaccurate
O
sensor not ready
2
Ventilation alarms
Alarm
Apnea (no flow)
Apnea (no pressure)
Apnea Ventilation
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Priority
Low
>20 s
2
Priority
Low
---
---
>20 s
Priority
Low
---
---
At the latest after 40 s in CPAP / PSV mode with
active RRmin.
(configurable alarm priority, see page 208)
Priority
Medium
---
measurement
Priority
Medium
After at least 5 s
---
Priority
Medium
At the latest after 20 s
(for RR ≥6)
or
At the latest after 35 s
(for RR <6 or in Manual /
Spontaneous mode)
At the latest after 20 s
(for RR ≥6)
or
At the latest after 35 s
(for RR <6)
Alarms
Priority
High
---
Priority
High
After at least 5 s
---
Priority
High
15 s later
(for RR ≥6)
or
30 s later
(for RR <6 or in Manual /
Spontaneous mode)
15 s later
(for RR ≥6)
or
30 s later
(for RR <6)
---
199