Dräger Atlan A100 Instructions For Use Manual page 282

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Technical data
For external breathing systems (restricted
range)
For internal breathing systems (restricted
range)
Connection for sample gas disposal when
using external patient-gas measurement
Outer hose diameter
Maximum inlet flow
15.12
Passive AGS
Information
Scavenging hose
Maximum length
Minimum diameter
Connection for sample gas disposal when
using external patient-gas measurement
Outer hose diameter
Maximum inlet flow
15.13
Measuring systems and displays
Information
Airway pressure Paw
Plateau pressure Pplat
Positive end-expiratory pressure PEEP
Peak inspiratory pressure PIP
Mean airway pressure Pmean
Range
Accuracy
Resolution of displayed value
Airway pressure gauge
Range
Accuracy
Resolution of displayed value
282
5 L/min
7 L/min
3 to 6 mm
500 mL/min
This system is designed for an anesthetic gas scav-
enging system which works without suction flow.
To limit contamination of the ambient air in compliance
with ISO 80601-2-13, follow the specifications for the
scavenging hose.
8 m (26 ft)
19 mm (0.75 in)
3 to 6 mm
500 mL/min
Alarms and internal calculations are based on values
that are more precise than the displayed values. Con-
sequently, there may be minor deviations between the
current alarm status and the displayed measured val-
ues.
–20 to +99 hPa (cmH
±4 % of the measured value or ±2 hPa (cmH
greater value applies)
1 hPa (cmH
–20 to 80 hPa (cmH
±5 % of the measured value or ±2 hPa (cmH
greater value applies)
5 hPa (cmH
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
O)
2
O)
2
O)
2
O)
2
O) (the
2
O) (the
2