Headline/
Parameter
Pressure
1 to 10
decrease per step
when PEEP ≤ 15
[mbar]; [hPa];
[cmH
O]
2
11.5.4.5
Vertical tab "Fresh gas" (only with electronically controlled gas mixer)
Headline/
Parameter
Start settings for fresh gas
FG O
21 to 100
2
[%]
FG flow
0.20 to
[L/min]
Minimal O
flow
Off; 50 to
2
(carrier gas: Air)
[mL/min]
Minimal O
flow
50 to 300
2
(carrier gas: N
O)
2
[mL/min]
Carrier gas
Air; N
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Setting range
1 to 10
2
2
Setting range
21 to 100
100
100
0.20 to
15.00
15.00
2.00
2.00
Off; 50 to
300
300
200
200
50 to 300
200
200
O
Air; N
O
2
2
1 to 10
For PEEP ≤15: Specifies the pressure
2
by which PEEP and Pinsp will be
reduced in steps.
Description
Selects the start settings for the fresh-
gas delivery.
21 to 100
Sets the O
100
gas.
0.20 to
Sets the fresh-gas flow.
15.00
2.00
Off; 50 to
Sets the minimal O
300
ered when Air is used as carrier gas.
200
Do not set the value too low, but suit-
able for the patient category. The rec-
ommendation is, e.g., 200 for adults,
100 for pediatric patients, and 50 for
neonates.
50 to 300
Sets the minimum O
200
ered when N
Do not set this value too small; recom-
mended is, e.g., 200 for adults, 100 for
pediatric patients, and 50 for neonates.
Air; N
O
Sets the carrier gas.
2
Configuration
Description
concentration in the fresh
2
flow that is deliv-
2
flow that is deliv-
2
O is used as carrier gas.
2
213