Dräger Atlan A100 Instructions For Use Manual page 286

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Technical data
Enflurane
Isoflurane
Sevoflurane
Desflurane
15.14
Display of calculated values
Information
Airway parameters
Dynamic compliance Cdyn
Mean dynamic compliance Cdyn mean
Range
Accuracy
Resolution of displayed value
Information
Resistance R
Range
Accuracy
Resolution of displayed value
Information
Elastance E
Range
Resolution of displayed value
Information
Time constant TC
Range
Resolution of displayed value
Information
Leakage
Difference between inspiratory and expira-
tory tidal volumes ∆VT
Range
Accuracy
Resolution of displayed value
286
0 to 6 Vol% (kPa)
0 to 5 Vol% (kPa)
0 to 10 Vol% (kPa)
0 to 20 Vol% (kPa)
The following calculated values are partly based on
measured values from various sensors with different
measurement accuracies. Due to the combination of
these measured values, the resulting accuracy of the
calculated values may be reduced.
0 to 200 mL/hPa (mL/cmH
±15 % of the displayed value or ±1 mL/hPa (cmH
(the greater value applies)
0.1 mL/hPa (mL/cmH
The values may be distorted by spontaneous breath-
ing, thus reducing the measurement accuracy.
0 to 100 hPa/L/s (cmH
±30 % of the displayed value or ±3 hPa/L/s
(cmH
O/L/s) (the greater value applies)
2
1 hPa/L/s (cmH
The values may be distorted by spontaneous breath-
ing, thus reducing the measurement accuracy.
0.005 to 10.000 hPa/mL (cmH
0.001 hPa/mL (cmH
The elastance E is calculated as 1 / Cdyn mean.
0.0 to 5.0 s
0.1 s
The time constant TC is calculated as R x Cdyn mean.
0 to 2500 mL
±20 % of the displayed value or ±30 mL (the greater
value applies)
1 mL
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
O)
2
O)
2
O/L/s)
2
O/L/s)
2
O/mL)
2
O/mL)
2
O)
2