10.3.2
Silencing the alarm tone
The alarm tone can be silenced for a maximum of 2 minutes.
● Press the alarm silence key
The
the header bar.
If the cause of the alarm persists, the alarm tone starts again immediately after the
alarm silence ends.
During the alarm silence, only new alarms with a higher alarm priority or a higher
internal priority number compared with the silenced alarm are acoustically signaled.
For further information see: "Alarm – Cause – Remedy", page 245.
10.3.2.1
Reactivating the alarm tone
● Press the alarm silence key
10.3.3
Downgrading and resetting alarm messages
Some alarms can be downgraded to low priority or they can be cleared completely.
The relevant alarms can be recognized in the table "Alarm – Cause – Remedy" on
page 245 by the following remedial messages:
Remedial message
Use "ALARM RESET" to downgrade
alarm priority.
Use "ALARM RESET" to reset alarm.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
symbol and the remaining time for the silenced alarm tone are displayed in
1
(1) below the screen, see page 21.
again.
Effect
Alarm priority is changed to low.
Alarm is cleared.
Alarms
193