Dräger Atlan A100 Instructions For Use Manual page 291

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Sensor sampling rate
Typical delay of the measured value
15.18
Operating characteristics
Mains power supply
Mains voltage
Maximum power consumption
Power cable
Maximum length
Protective ground resistance
Operating voltage
Operating current
Current consumption at 230 V AC
When device is switched off with switched-on
main switch and fully charged battery
During mechanical ventilation (PC - CMV,
Pinsp = 15 hPa, RR = 8 /min, I:E = 1:2,
PEEP = 0 hPa, FG flow = 4 L/min O
charging the internal battery
Maximum
Current consumption at 110 V AC
When device is switched off with switched-on
main switch and fully charged battery
During mechanical ventilation (PC - CMV,
Pinsp = 15 hPa, RR = 8 /min, I:E = 1:2,
PEEP = 0 hPa, FG flow = 4 L/min O
charging the internal battery
Maximum
Power consumption
When device is switched off with switched-on
main switch and fully charged battery
During mechanical ventilation (PC - CMV,
Pinsp = 15 hPa, RR = 8 /min, I:E = 1:2,
PEEP = 0 hPa, FG flow = 4 L/min O
charging the internal battery
Maximum
Peak inrush current
Internal battery
Type
Nominal capacity
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
<40 ms
<15 s
100 to 240 V AC at 50/60 Hz
4 A
5 m (16.4 ft)
Maximum 0.1 Ω
≥250 V
≥10 A
<0.18 A
<0.40 A
) without
2
2 A
<0.38 A
<0.84 A
) without
2
4 A
<40 W
<95 W
) without
2
400 W
Approx. 8 to 14 A
Approx. 6 to 10 A quasi-RMS
Lead battery
Sealed, maintenance-free
7.2 Ah
Technical data
291