Dräger Atlan A100 Instructions For Use Manual page 209

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

11.5.3.5
Vertical tab "Config. 2"
Headline/
Parameter
Deactivate the alarm limit in the following modes: Man/Spon,
Pause, CPAP/PSV with ∆Psupp < 5, ext. FGO
Deactivate the alarm limit in the following modes: Man/Spon,
CPAP/PSV with ∆Psupp < 5, ext. FGO, Monitoring
FiO
high
Yes; No
2
MV low
MV high
xMAC low
etCO
low
2
etCO
high
2
inCO
high
2
inAgent low
Deactivate the alarm limit in cardiac bypass mode (CBM)?
FiO
high
2
MV low
MV high
inAgent low
1) Pause must be activated in the system setup.
2) Only for devices with integrated PGM. Pause must be deactivated in the system setup.
11.5.3.6
Vertical tab "Config. 3"
Headline/
Parameter
Alarm limits for "Cylinder almost empty"
O
2
Air
N
O
2
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Setting range
1)
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Yes; No
Setting range
[bar]; [kPax100]: Off; 15 to 50
20
[psi]: Off; 218 to 725
290
[bar]; [kPax100]: Off; 15 to 50
20
[psi]: Off; 218 to 725
290
[bar]; [kPax100]: Off; 15 to 40
20
[psi]: Off; 218 to 580
290
Description
2)
Yes; No
Specifies the alarm behavior when
changing to a different ventilation
mode.
These settings apply only when switch-
ing to a ventilation mode with lower or
no breathing support (see page 310).
The alarm behavior at the start of the
therapy is defined by the configuration
in the vertical Limits tab.
Specifies the alarm behavior in the
CBM mode.
Description
Specifies the alarm limits for the supply
pressure of connected gas cylinders.
Configuration
209