Dräger Atlan A100 Instructions For Use Manual page 87

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

4.8.2
Permissible hose configurations
Breathing hoses, sample line, filters, etc. must be arranged carefully and
adapted to the patient, particularly for neonates and pediatric patients. For further
information see: "CO
page 321.
The permissible hose configurations are marked with an "X" in the following table:
Configuration
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
measurement with pediatric patients and neonates",
2
A breathing system filter or an HME filter
between Y-piece and patient; sample line
connected to the filter
One breathing system filter on each of
the inspiratory port and expiratory port;
sample line connected to the Y-piece
– One breathing system filter each on
the inspiratory port and expiratory
port; sample line connected to a con-
nector as close as possible to the
patient
– Extract the sample gas at a suitable
connector.
– If no filter can be used at the expira-
tory port (e.g., if there is intrinsic
PEEP due to Air trapping), reprocess
the breathing system after every
patient.
Perform the reprocessing in accor-
dance with the reprocessing instruc-
tions supplied with the product.
Assembly and preparation
Adult
Ped.
Neo.
X
X
X
X
X
X
X
87