Dräger Atlan A100 Instructions For Use Manual page 207

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Headline/
Parameter
0.10 to 8.50
inIso
[%]; [kPa]
inIso
0.00 to 8.40
[%]; [kPa]
0.10 to 8.50
inHal
[%]; [kPa]
inHal
0.00 to 8.40
[%]; [kPa]
11.5.3.2
Vertical tab "Sound volume"
Headline/
Parameter
Alarm volume
Minimum alarm
volume
11.5.3.3
Vertical tab "Autoset"
Headline/
Parameter
Offset for "Autoset limits" function
Automatic adjustment of the parameters to current measured values. For further
information, see the following section: "Automatic setting", page 179..
By touching the Autoset button, the alarm limits are adjusted so that the upper
alarm limit is above the current measured value by at least the percentage or value
set here and the lower alarm value is correspondingly below it.
Example:
In the PC - CMV mode:
Measured MV: 5 L/min
Set offset: ±40 %
New alarm limits: 7 and 3 L/min
etCO
± [%]
2
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Setting range
0.10 to 8.50
2.40
2.80
Off;
Off;
0.00 to 8.40
0.10 to 8.50
1.60
1.90
Off;
Off;
0.00 to 8.40
Setting range
10 to 100
40
10 to 100
10
Setting range
Off; 20 to 80
20
Description
0.10 to 8.50
Isoflurane
3.70
Off;
0.00 to 8.40
0.10 to 8.50
Halothane
2.40
Off;
0.00 to 8.40
Description
Sets the alarm volume.
Sets the minimum volume with which
an alarm tone will be signaled.
Description
In modes with low or no breathing sup-
port (Man/Spon, Ext. FGO, CPAP /
PSV, Pause, and Monitoring), a further
20 percentage points are added to the
configured value.
Configuration
207