12.8.1.2
Removing the flow sensors and the ports
1. Unfasten and remove the expiratory port (1).
2. Remove the expiratory flow sensor (2).
3. Unfasten and remove the inspiratory port (3).
4. Remove the inspiratory flow sensor (4).
12.8.1.3
Fitting the flow sensors and the ports
1. Insert the expiratory flow sensor (1).
2. Push in the expiratory port (2). Tighten the knurled nut.
3. Insert the inspiratory flow sensor (3).
4. Push in the inspiratory port (4). Tighten the knurled nut.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n