Priority
Alarm
!! 50
Fresh-gas flow inaccurate The accuracy of the fresh-
!! 255
Gas sensor failure
!! 255
Hose connected to wrong
port
!! 100
Hose does not fit to pat.
category
!
100
Hose does not fit to pat.
category
!! 150
inCO
high
2
!
100
Infinity ID breathing circuit
not compatible
!
100
Infinity ID CO
not compatible
!
100
Infinity ID water trap not
compatible
!!! 255
Inspiratory desflurane high Inspiratory anesthetic gas
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Cause
gas flow measurement is
reduced.
The patient-gas measure-
ment module has failed.
A breathing hose is not cor-
rectly connected.
The detected breathing hose
is not suitable for the
selected patient category.
The detected breathing hose
is not suitable for the
selected patient category.
Soda lime is depleted.
There is leakage in the
breathing system or in the
coaxial breathing hose.
Gas measurement is inaccu-
rate due to high respiratory
rate.
Large dead space.
An incompatible accessory is
connected.
absorber
An incompatible accessory is
2
connected.
An incompatible accessory is
connected.
concentration has exceeded
the upper alarm limit.
Troubleshooting
Remedy
Use the total flow tube to ver-
ify the current fresh-gas flow.
Check the measured gas
concentrations.
Use "ALARM RESET" to
downgrade alarm priority.
Perform the system test.
If the problem persists, call
Dräger.
Use an alternative gas mea-
surement system.
Use "ALARM RESET" to
downgrade alarm priority.
Call Dräger.
Connect breathing hoses cor-
rectly.
Use compatible accessory.
Use "ALARM RESET" to
reset alarm.
Use compatible accessory.
Use "ALARM RESET" to
reset alarm.
Check soda lime.
Increase fresh-gas flow.
Check fresh-gas settings.
Replace the breathing sys-
tem or the coaxial breathing
hose.
Adjust alarm limits if neces-
sary.
Check the ventilation settings
and the breathing circuit.
Check accessory.
Use "ALARM RESET" to
reset alarm.
Check accessory.
Use "ALARM RESET" to
reset alarm.
Check accessory.
Use "ALARM RESET" to
reset alarm.
Check vaporizer and fresh-
gas settings.
255