16.2.5.4
VC - SIMV
– Volume-controlled
– Pressure-limited
– Time-controlled
– Machine-triggered or patient-triggered
– Constant inspiratory flow
– Synchronized inspiration
In VC - SIMV, the patient can breathe spontaneously at any time, while the number
of mandatory breaths is predefined. When synchronization is switched on, the
breaths are adapted to the spontaneous breathing of the patient. If inspiratory effort
by the patient is detected during the inspiratory trigger window, a patient-triggered
breath will be initiated.
16.2.5.5
VC - SIMV / PS
This mode is similar to VC - SIMV, except that the patient's spontaneous breathing
at the PEEP level during the expiratory phase is pressure-supported with ∆Psupp
when outside the trigger window.
16.2.5.6
VC - AF
For volume-controlled mandatory breaths with AutoFlow, the set tidal volume VT is
applied with the lowest required pressure. The pressure patterns and flow patterns
of the mechanical inspiratory breaths correspond to those of pressure-controlled
ventilation.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Paw
Pmax
Slope
Ti
Flow
PEEP
∆Psupp
Slope
1/RR
Trigger
Insp term
Principles of operation
Trigger
315