Strangulation
Negligent placement of hoses, cables, and similar device components can put the
patient at risk.
► Use particular caution when establishing connections to the patient.
Transport
If the device collides with an obstacle during transport, the pressure reducers may
be damaged. This may cause a fire. Take the following measures before
transporting:
► Align the pressure reducers so that they are protected from collisions.
► Close the valves on the gas cylinders.
► If no gas cylinders are connected, fasten the pressure reducers with the hook-
and-loop straps of the cylinder holders.
During transport, the device may tip over due to incorrect handling or carelessness.
Personal injury or property damage may result as a consequence.
► Remove accessories mounted on the column cover before transport.
► When pushing the device, hold on to the standard rail with handle (right side of
the device).
► The device may only be moved by persons who have the physical ability to do
so.
► Always have the device moved by 2 persons for better maneuverability and
when it is being transported over sloping surfaces.
► When the device is moved over inclines, around corners, or over thresholds
(e.g., through doors or in elevators), make sure that it does not bump against
anything.
► Do not pull the device over hoses, cables, or other obstacles lying on the floor.
► Do not operate either the central brake or any of the castor brakes while the
device is being moved.
► Do not lean against the device.
► After transport, perform a visual inspection for damage.
4.1.7
Color codes and labels
If the vaporizer is filled with the wrong anesthetic agent or if the filling level is too
low, incorrect anesthetic gas concentrations or concentrations that are too low can
occur as a result.
► Compare the color code and labeling on the vaporizer used with the anesthetic
agent bottle and the anesthetic agent indicated on the screen.
► Observe the vaporizer filling level.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Assembly and preparation
65