Dräger Atlan A100 Instructions For Use Manual page 211

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

11.5.4.2
Vertical tab "Vent. 2"
Headline/
Parameter
Start settings for ventilation
Pmax
12 to 80
[mbar]; [hPa];
[cmH
O]
2
Pinsp
7 to 80
[mbar]; [hPa];
[cmH
O]
2
∆Psupp
[mbar]; [hPa];
3 to 80
[cmH
O]
2
Insp term
5 to 80
[%]
PEEP
[mbar]; [hPa];
2 to 35
[cmH
O]
2
Slope
[s]
RRmin
[/min]
3 to 25
I:E
1:10 to 4:1
% Tplat
0 to 60
[%]
Trigger
0.3 to 15
[L/min]
Sync.
SIMV; CMV SIMV; CMV SIMV; CMV
11.5.4.3
Vertical tab "Proced. 1"
Headline/
Parameter
General settings
Default maneuver One-step recruitment; Multi-step recruit-
Reminder
Off, 10 to
[min]
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Setting range
12 to 80
40
30
7 to 80
15
15
Off;
Off;
3 to 80
10
10
5 to 80
25
25
Off;
Off;
2 to 35
3
3
0 to 2
0 to 1.5
0.2
0.2
Off;
Off;
3 to 25
6
10
1:10 to 4:1
1:2
1:2
0 to 60
20
20
0.3 to 15
4.0
2.0
Setting range
ment; Insp./Exp. hold
Off, 10 to
180
180
Description
Specifies the start settings for the venti-
lation.
12 to 80
25
7 to 80
15
Off;
3 to 80
10
5 to 80
25
Off;
2 to 35
3
0 to 1.5
0.2
Off;
3 to 25
15
1:10 to 4:1
1:1
0 to 60
20
0.3 to 15
1.0
Defines the default maneuver.
Off, 10 to
If the value is not set to Off, a reminder
180
for a maneuver is given after the first
switch to a controlled ventilation mode.
Defines the time after which a reminder
for a further maneuver is issued after
the end of a One-step recruitment or
Multi-step recruitment maneuver.
Configuration
Description
211