Operation
8.6.7
MV×CO
Prerequisite: The device has the "advanced gas monitoring" software option.
MV×CO2
The MV×CO
calculated as the product of the minute volume (MV) and the difference of the
measured expiratory CO
concentrations (inCO
8.6.8
ΔVT
ΔVT
VT
Paw
The ∆VT parameter field displays the inspiratory tidal volume (VTi) and expiratory
tidal volume (VT) measured values and the difference between these two
measured values (∆VT). In addition, the expiratory tidal volume (VT) and airway
pressure (Paw) measured values are displayed as bar graphs.
8.6.9
Compliance trend
Prerequisite: The device has the "advanced ventilation monitoring" or "loops and
trends" software option.
Cdyn
The Compliance trend parameter field displays the current measured values for the
dynamic patient compliance (Cdyn and PEEP) as well as the trend of the mean
dynamic patient compliance (Cdyn mean).
168
trend
2
Now
min
min
parameter field displays the CO
2
concentrations (etCO
2
).
2
VTi
PEEP
Now
min
min
Instructions for use
elimination trend. The trend is
2
) and inspiratory CO
2
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n