Anesthesia Gas Supply Module
Fresh-Gas Metering, Low-Flow Flow Meters
(calibrated vor 68 °F (20 °C), 1013 cmH
O
2
N
O
2
AIR
O
Flush (bypass):
2
Common Gas Outlet Pressure Limit:
Auxiliary Oxygen Flow Meter (optional)
Connection
Fresh-gas flow
Accuracy
Resolution
Instructions for Use Fabius plus SW 3.n
O (hPa)):
2
0.02 to 10.0 L/min: ±10 % of the indicated value for flows
between 10 % and 100 % of full scale
according to IEC 60601-2-13 ambient atmosphere of 14,7 psi
(101,3 kPa) at an operating temperature of 20 °C (± 3 °C).
0.02 to 10.0 L/min: ±10 % of the indicated value for flows
between 10 % and 100 % of full scale
according to IEC 60601-2-13 ambient atmosphere of 14,7 psi
(101,3 kPa) at an operating temperature of 20 °C (± 3 °C).
0.02 to 12.0 L/min: ±10 % of the indicated value for flows
between 10 % and 100 % of full scale
according to IEC 60601-2-13 ambient atmosphere of 14,7 psi
(101,3 kPa) at an operating temperature of 20 °C (± 3 °C).
at 87 psi (6 kPa x 100): max. 75 L/min
at 41 psi (2.8 kPa x 100): min. 25 L/min
13 psi (0.9 kPa x 100), maximum
Staged connector for use with different hose diameters
0 to 10 L/min
±5 % of full scale
0.5 L/min
Technical Data
203