Dräger Atlan A100 Instructions For Use Manual page 263

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Priority
Alarm
!! 255
O
sensor failure
2
!! 255
O
sensor not ready
2
!!! 210
O
supply low
2
!!! 210
O
supply low
2
!
100
On/Standby key stuck
!
75
Patient-gas measurement
module is calibrating
!!! 230
Pause time expired
!! 230
Pause time expired
!! 100
PEEP/CPAP low
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Cause
The O
sensor is faulty.
2
A new O
sensor cell was
2
inserted.
The O
sensor calibration is
2
invalid.
Central supply pressure and
cylinder pressure are low.
Central supply pressure and
cylinder pressure are low.
Key is stuck or was pressed
for more than 10 seconds.
The automatic calibration of
the patient-gas measurement
module is in progress.
Ventilation and gas delivery
have been paused longer
than the set pause time.
Ventilation and gas delivery
have been paused longer
than the set pause time.
Fresh-gas flow is insuffi-
cient, breathing bag is
blocked or positioned incor-
rectly.
Leakage or disconnection.
Failure of the anesthetic gas
scavenging system.
Troubleshooting
Remedy
Use substitute O
monitoring.
2
Use "ALARM RESET" to
downgrade alarm priority.
When the case has been
completed, perform a sys-
tem test.
Replace the sensor cell if
necessary.
If the problem persists, call
Dräger.
Wait for the O
sensor to
2
warm up.
Perform a system test and
calibrate the O
sensor.
2
Use "ALARM RESET" to
downgrade alarm priority.
Perform a system test and
calibrate the O
sensor.
2
Check central O
supply or
2
use cylinder.
Use "ALARM RESET" to
downgrade alarm priority.
Check central O
supply or
2
use cylinder.
Use "ALARM RESET" to
downgrade alarm priority.
The therapy will continue with
current settings.
If the problem persists, call
Dräger.
Wait for the calibration to
complete.
Resume ventilation or adjust
timer.
Resume ventilation or adjust
timer.
Check fresh-gas settings and
position of breathing bag.
Check the breathing circuit
for tight connections and
leakages.
Check anesthetic gas scav-
enging system.
263