Dräger Atlan A100 Instructions For Use Manual page 295

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Transmitter power
Modulation
Electromagnetic compatibility
Protection classes
Device
Applied parts (connections for breathing
hoses)
Degree of protection
Classification Medical Device Europe
UMDNS code
Use of latex
15.19
Interfaces and ports
Serial ports
Information
Protocol
Alarm delay time (measured as of the query
time)
Connector
Baud rate
Data bits
Parity
Stop bits
Pin assignment
Pin 1
Pin 2
Pin 3
Pins 4, 6
Pin 5
Pins 7, 8
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
≤42 dBμA/m (200 mW ±1 dB)
ASK (amplitude shift keying)
Tested in compliance with IEC 60601-1-2
I, in compliance with IEC 60601-1
TYPE BF
IP11 against vertically falling dripping water under typi-
cal conditions of use in accordance with the intended
use (with 0° inclination of the device)
IP10 at >0° and ≤5° inclination of the device
In accordance with IEC 60529, meets ISO 80601-2-13
Class II b
10-134
The device is made without natural rubber latex.
COM 1 and COM 2
Only connect devices that meet the requirements of
IEC 60950-1 / IEC 62368-1 for ungrounded SELV cir-
cuits or the requirements of IEC 60601-1 (as of the 2nd
edition) for touchable secondary circuits with a maxi-
mum nominal voltage of 24 V DC.
Use COM 2 to connect devices for external patient-gas
measurement with the MEDIBUS communication pro-
tocol.
MEDIBUS.X
<3 s for baud rates ≥9600
<10 s for baud rates <9600
9-way Sub-D
1200, 2400, 4800, 9600, 19200, 38400 baud
8
Even
1
Not used
RXD
TXD
Pins 4 and 6 are connected internally
SHLD-GND
Pins 7 and 8 are connected internally
Technical data
295