Advertisement

5. Check the safety equipment:
– Functional integrity of optical and acoustic alarm generators
– O
– Internal battery
– Check the function of the minimum O
6. Check the accuracy of the gas measurement of the patient-gas measurement
module based on a certified test gas concentration:
– Anesthetic gas measurement:
– N
– CO
7. Check the accuracy of the O
– Ambient air 21 Vol%
– 100 Vol%
8. Check sample gas flow of the patient-gas measurement module:
– Accuracy 200 ±20 mL/min
9. Check the patient-gas measurement module for leakage:
– Leakage at –200 hPa (cmH
– <20 hPa/min (cmH
10.Check the non-return valves in the central supply for leakage:
– Leakage ≤20 mL/min
11. Check the operational readiness by means of a system test.
12.Check the accuracy of the airway pressure gauge:
– Accuracy 30 hPa (cmH
13.Perform a visual inspection of both safety valves of the passive AGS for damage
and soiling.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
switch on the electronically controlled gas mixer
2
3-gas gas mixer).
Isoflurane, 1 Vol%
Sevoflurane, 1 Vol%
Accuracy ±0.35 Vol%
O measurement, 70 Vol%
2
Accuracy ±7.6 Vol%
measurement, 5 Vol%
2
Accuracy ±0.83 Vol%
Accuracy ±3 Vol%
Accuracy –5 Vol%
O/min)
2
delivery (mechanically controlled
2
measurement:
2
O)
2
O) ±10 hPa (cmH
O)
2
2
Service
269