Assembly and preparation
4.9.2.2
Filling and mounting
1. Push the absorber insert (5) into the absorber container (6).
CAUTION
Risk of chemical burns
Soda lime is caustic and is a strong irritant for eyes, skin, and airway.
► Handle the soda lime carefully and do not spill it.
WARNING
Risk of carbon monoxide formation
If soda lime based on potassium hydroxide is used, there is a risk of carbon
monoxide formation.
► Do not use soda lime based on potassium hydroxide.
► Use a suitable soda lime such as Drägersorb Free.
2. Fill the CO
Recommendation: Use Drägersorb 800 Plus or Drägersorb Free.
WARNING
Risk of hypoventilation and incorrect gas measurement
Reuse of the disposable dust filter can increase the filter resistance and impair the
ventilation function of the device.
► If the reusable CO
Replace the disposable dust filter with every change of soda lime.
3. If soda lime from third-party manufacturers is used, insert a new disposable dust
filter (7). Only use dust filters from the list of accessories.
Only use undamaged filters.
4. Attach the CO
direction of the arrow until it reaches the stop.
94
5
6
absorber with fresh soda lime to the upper mark.
2
absorber is used, always use a disposable dust filter.
2
absorber (8) to the breathing system from below. Rotate it in the
2
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n