Technical data
15.15
Gas measurement
Information
15.16
Gas measurement with PGM
Information
Standardization
Accuracies
Sample gas flow
Maximum time until emptying of the water
trap is necessary
System response time
Sensor sampling rate
Time after switch-on until the specified
accuracy is attained
Time until measured CO
played (with at least reduced accuracy)
Typical delay by the sample line
Cross sensitivity
Drift
Calibration
O
2
Measuring principle
Range
Accuracy
Resolution of displayed value
T
time of the sensor
10...90
288
The device is equipped with a patient-gas measure-
ment module (PGM) for O
agent, or with an O
This is a sidestream gas measurement, in which the
sample gas is fed back into the breathing system.
End-tidal measured values are calculated for each
breath from the local maxima and minima of the real-
time measurements during expiration.
All data are standardized to ATPS conditions.
Due to the T
measurement accuracies for O
thetic agent may deviate at respiratory rates of 60 /min
or higher and an I:E ratio of 1:1. The influence of respi-
ratory rate and the I:E ratio on the accuracy has been
verified in a simulated breathing system using a rectan-
gular waveform for the gas concentration.
200 ±20 mL/min
standardized to STPD conditions
41 hours (sample gas under BTPS conditions and at
23 °C (73.4 °F) ambient temperature)
The system response time results from the typical
delay by the sample line and the gas-type-specific
T
10...90
<50 ms
<270 s at 20 °C (68 °F) ambient temperature
<480 s at 10 °C (50 °F) ambient temperature
values are dis-
<90 s
2
<2.5 s with Dräger sample line, 2.5 m (98.4 in)
None concerning alcohol (<3000 ppm in blood), ace-
tone (<1000 ppm), methane, water vapor, NO, and CO
Compensated by automatic cyclic calibration
During automatic calibration, ambient air is fed to the
breathing system and used as the sample gas.
Paramagnetic (consumption-free)
0 to 100 Vol%
±(2.5 Vol% + 2.5 % of the measured value)
1 Vol%
<500 ms
Instructions for use
, CO
2
sensor.
2
time and the sample gas flow, the
10...90
time of the sensor.
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
, N
O, and anesthetic
2
2
, CO
, N
O, and anes-
2
2
2