Dräger Atlan A100 Instructions For Use Manual page 215

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Headline/
Parameter
I:E changes with
RR in SIMV
Breathing system
warmer
Auto Wake-up
(opens start dia-
log upon respira-
tory activity)
Anesthetic gas for
flow correction
Auto
None
Sev
Des
Other
Carrier gas for
flow correction
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Setting range
On; Off
On; Off
On; Off
Auto; Des; Sev;
Other; None
0 to 6.5
2.1
0 to 18
6
Auto; Air; N
O
2
Description
Determines whether the change of RR
automatically causes a change of I:E. In
this way, Ti remains constant.
Applies to the following ventilation
modes:
– PC - SIMV
– PC - SIMV / PS
– VC - SIMV
– VC - SIMV / PS
– VC - SIMV / AutoFlow
– VC - SIMV / PS / AutoFlow
Prerequisite: Inspiratory-time control is
set to I:E.
Switches the breathing system warmer
on or off.
Determines whether the Start dialog
opens automatically when ventilation
activity is detected (e.g., as a result of
repeated squeezing of the breathing
bag).
Corrects the accuracy of the flow mea-
surement on devices with inspiratory
O
measurement.
2
Corrects the measured flow values
automatically when an anesthetic gas
monitor is connected (see "Flow correc-
tion", page 149).
Deactivates the correction.
Corrects the measured flow values
when sevoflurane is used.
Select the value according to the typical
setting on the vaporizer control dial.
(The vaporizer automatically makes
any correction necessary for altitude.)
Corrects the measured flow values
when desflurane is used.
Select the value according to the typical
setting on the vaporizer control dial.
(The vaporizer automatically makes
any correction necessary for altitude.)
Deactivates the correction.
Corrects the accuracy of the flow mea-
surement on devices with inspiratory
O
measurement and mechanically
2
controlled gas mixer with flow tubes.
Configuration
215