Dräger Atlan A100 Instructions For Use Manual page 278

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Technical data
Peak inspiratory flow
Pressure-related settings
Inspiratory pressure Pinsp
Setting range
Increment value
Accuracy
Pressure limitation Pmax
Setting range
Increment value
Accuracy
Relative pressure support above PEEP
∆Psupp in CPAP / PSV and PC - SIMV / PS
modes
Setting range
Increment value
Accuracy
Relative pressure support above PEEP
∆Psupp in VC - SIMV / PS and VC - SIMV /
PS / AutoFlow modes
Setting range
Increment value
Accuracy
Positive end-expiratory pressure PEEP
Setting range
Increment value
Accuracy
278
180 L/min or higher
Results from the following set values:
- VT and Ti in volume-controlled ventilation modes
- Pinsp and Ti in pressure-controlled ventilation modes
Applies for mains operation or in the first 5 minutes of
battery operation with a fully charged battery. In other
situations, the peak inspiratory flow may be restricted
to 75 L/min.
PEEP +5 to 80 hPa (cmH
(7 to 80 hPa (cmH
1 hPa (cmH
±10 % of the set value or ±3 hPa (cmH
value applies)
PEEP +5 to 80 hPa (cmH
(7 to 80 hPa (cmH
1 hPa (cmH
±10 % of the set value or ±3 hPa (cmH
value applies)
Off, 3 to (80 - PEEP) hPa (cmH
(Off, 3 to 78 hPa (cmH
1 hPa (cmH
±10 % of the set value or ±3 hPa (cmH
value applies)
Off, 3 to (Pmax - PEEP) hPa (cmH
(Off, 3 to (Pmax - 2) hPa (cmH
1 hPa (cmH
±10 % of the set value or ±3 hPa (cmH
value applies)
Off, 2 to 35 hPa (cmH
1 hPa (cmH
±10 % of the set value or ±2 hPa (cmH
value applies)
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
O)
2
O) when PEEP = Off)
2
O)
2
O)
2
O) when PEEP = Off)
2
O)
2
O)
2
O) when PEEP = Off)
2
O)
2
2
O) when PEEP = Off)
2
O)
2
O)
2
O)
2
O) (the greater
2
O) (the greater
2
O) (the greater
2
O)
O) (the greater
2
O) (the greater
2