Dräger Atlan A100 Instructions For Use Manual page 287

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Leakage minute volume MVleak
Range
Accuracy
Resolution of displayed value
Measurement of consumption and
elimination
CO
elimination per minute
2
Standardization
Range
Accuracy
Resolution of displayed value
O
uptake per minute
2
Standardization
Range
Accuracy
Resolution of displayed value
Anesthetic agent uptake
Range
Accuracy
Resolution of displayed value
Anesthetic agent consumption
Range
Accuracy
Resolution of displayed value
Fresh-gas consumption
Standardization
Range
Accuracy
Resolution of displayed value
Information
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
0.00 to 40.0 L/min
±25 % of the displayed value or ±0.2 mL/min (the
greater value applies)
0.01 L/min (with MVleak <1 L/min)
0.1 L/min (with MVleak ≥1 L/min)
The following data are standardized to STPD condi-
tions.
0 to 9999 mL/min
±25 % of the displayed value or ±100 mL/min (the
greater value applies)
1 mL/min
The following data are standardized to STPD condi-
tions.
0 to 9999 mL/min
±25 % of the displayed value or ±100 mL/min (the
greater value applies)
1 mL/min
0.0 to 99.9 mL fluid
±25 % of the displayed value or ±1 mL (the greater
value applies)
0.1 mL
0.0 to 999.9 mL fluid
±25 % of the displayed value or ±1 mL (the greater
value applies)
0.1 mL
The following data are standardized to STPD condi-
tions.
0 to 99999 L
±15 % of the displayed value or ±2 L/min (the greater
value applies)
1 L
Gas consumption related to O
not included in this calculation.
Technical data
flush and O
therapy is
2
2
287