Troubleshooting
Priority
Alarm
!! 255
Inspiratory desflurane high Inspiratory anesthetic gas
!
255
Inspiratory desflurane low The inspiratory anesthetic
!!! 255
Inspiratory enflurane high
!! 255
Inspiratory enflurane high
!
255
Inspiratory enflurane low
!
100
Inspiratory flow sensor
expired
!
255
Inspiratory flow sensor
failure
!!! 255
Inspiratory halothane high Inspiratory anesthetic gas
!! 255
Inspiratory halothane high Inspiratory anesthetic gas
!
255
Inspiratory halothane low
!!! 255
Inspiratory isoflurane high Inspiratory anesthetic gas
!! 255
Inspiratory isoflurane high Inspiratory anesthetic gas
256
Cause
concentration has exceeded
the upper alarm limit.
gas concentration is below
the lower alarm limit.
The soda lime has dried out. Replace the soda lime.
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
The inspiratory anesthetic
gas concentration is below
the lower alarm limit.
The soda lime has dried out. Replace the soda lime.
Accessory has been used
too long.
Failure of the flow sensor.
concentration has exceeded
the upper alarm limit.
concentration has exceeded
the upper alarm limit.
The inspiratory anesthetic
gas concentration is below
the lower alarm limit.
The soda lime has dried out. Replace the soda lime.
concentration has exceeded
the upper alarm limit.
concentration has exceeded
the upper alarm limit.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Remedy
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Refill the vaporizer.
Check the breathing system
for high leakage.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Refill the vaporizer.
Check the breathing system
for high leakage.
Replace the accessory if nec-
essary.
Use "ALARM RESET" to
reset alarm.
Replace the flow sensor and
perform a leakage test.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Refill the vaporizer.
Check the breathing system
for high leakage.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.