Dräger Atlan A100 Instructions For Use Manual page 258

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Troubleshooting
Priority
Alarm
!! 100
Internal device tempera-
ture high
!
255
Internal device tempera-
ture high
!!! 150
Internal FG flow failure
!
150
Internal FG flow failure
!
60
License expired
!
50
License will expire soon
258
Cause
A ventilation slot at the rear
of the device is blocked.
Ambient temperature is too
high.
A fan is faulty.
Excessive ventilation set-
tings are applied (e.g., high
respiratory rate, high inspira-
tory pressure, short slopes).
A ventilation slot at the rear
of the device is blocked.
Ambient temperature is too
high.
Excessive ventilation set-
tings are applied (e.g., high
respiratory rate, high inspira-
tory pressure, short slopes).
A fan is faulty.
The internal gas delivery sys-
tem is not operational. A sys-
tem test may be able to
resolve the issue.
The internal gas delivery sys-
tem is not operational. A sys-
tem test may be able to
resolve the issue.
A license has expired. After
next startup, some functions
will no longer be available.
A trial license will expire
within the next 14 days.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Remedy
Check the ventilation slots.
Ensure air flow at the rear of
the device.
Check ambient conditions.
Call Dräger.
Check ventilation settings.
Check the ventilation slots.
Ensure air flow at the rear of
the device.
Check ambient conditions.
Check ventilation settings.
Call Dräger.
Deliver O
: 1. Set the O
2
switch to "Add. O
". 2. Set
2
the O
flowmeter to the
2
desired flow. Check the
vaporizer setting. Make sure
that fresh gas is reaching the
patient.
When the case has been
completed, perform a sys-
tem test.
If the problem persists, call
Dräger.
Set the O
flowmeter to the
2
desired flow. Check the
vaporizer setting. Make sure
that fresh gas is reaching the
patient.
When the case has been
completed, perform a sys-
tem test.
If the problem persists, call
Dräger.
To order a permanent
license, call Dräger.
Use "ALARM RESET" to
reset alarm.
To order a permanent
license, call Dräger.