7.5.6.5
Calibrating the inspiratory O
During calibration, the sensor must be exposed to ambient air, i.e., 21 % oxygen
concentration. Fluctuating oxygen concentrations must be avoided. The following
notes must be adhered to:
– Do not fill vaporizers during the calibration.
– Close all the flow control valves.
– Do not blow into the sensor.
– Do not use any disinfectants.
1. If the O
Calibrate button.
If the validity of the last calibration has expired, the prompt to perform a
calibration appears automatically.
2. Touch the
3. Turn the O
Place the O
4. Take the sealing cap (2) from the holder (3).
Seal the sensor port with the sealing cap.
5. Follow the instructions on the screen.
The device will perform the calibration.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
sensor
2
sensor is to be calibrated despite a valid calibration, touch the
2
button and follow the instructions on the screen.
sensor counterclockwise and remove it.
2
sensor as shown (1).
2
1
Tests
137