Operation
8.5.5
Alarm behavior during the maneuvers
During the maneuvers, some alarms are adapted as follows:
Alarm
Man. insp./
Pressure alarms:
Airway pressure
high
Airway press.
continuously high
Inspiratory pres-
sure not
Switched off
achieved
Volume alarms:
Minute volume
Is suppressed during the maneuver and for an additional
low
Apnea alarms:
Apnea (no pres-
sure)
Apnea (no flow)
Apnea (no CO
)
2
1) Existing alarms are preserved. No new alarms will be issued.
Some technical alarms stop the maneuver automatically, e.g., alarms caused by a
sensor failure.
162
Insp./Exp. hold
Exp. hold
Insp. hold
---
1)
Stopped
Switched off
45 seconds after the end of the maneuver.
Switched off
Stopped
Instructions for use
Maneuvers
One-step
recruitment
Alarm limit is
---
adjusted to
Pressure + 5
Alarm limit is
---
adjusted to
Pressure + 3
Switched off
1)
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Multi-step
recruitment
Alarm limit is adjusted
to Pinsp max + 5
Alarm limit is adjusted
to half way between
PEEP max and Pinsp
max.
Example:
PEEP max = 20 hPa
Pinsp max = 40 hPa
Adjusted alarm limit =
30 hPa
---
---
---