Principles of operation
16.2.5
Volume-controlled ventilation
16.2.5.1
Compliance compensation
The applied VT is corrected by the determined breathing hose compliance, i.e., an
additional volume is delivered in order to ensure the application of the volume to the
patient. If the device is equipped with the integrated patient-gas measurement
module, the applied VT is corrected by the amount of suction flow as soon as CO
respiratory phases are detected.
16.2.5.2
VC
16.2.5.3
VC - CMV
– Volume-controlled
– Pressure-limited
– Time-controlled
– Machine-triggered
– Constant inspiratory flow
In this volume-controlled ventilation mode, the patient receives the set tidal volume
VT with each mandatory breath.
314
Paw
% Tplat
Ti
Flow
Instructions for use
∆Psupp
Slope
1/RR
Trigger
Insp term
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Pmax
PEEP
Trigger
2