Dräger Atlan A100 Instructions For Use Manual page 49

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

3.2.11
Data exchange, interfaces
3.2.11.1
Serial port
Two serial ports, COM 1 and COM 2, are provided for data transmission using the
MEDIBUS.X communication protocol.
3.2.11.2
USB port
After a suitable USB flash drive is connected, the USB port enables, e.g., the
following actions:
– Saving the screen contents as a screenshot.
– Saving and loading device configurations.
– Saving system test results or records as a text file.
Note the additional information on the specification of the USB port (see "Technical
data", page 272).
3.2.11.3
Network port
If an appropriate service contract has been obtained, the Dräger Remote Service
function can be executed.
The device can be connected to the ServiceConnect Gateway or a service laptop.
If the connected network offers an NTP service, the time on the device can be
synchronized with the time on the NTP server.
For further information see: "IT networks and cybersecurity", page 301.
3.2.11.4
Support of Infinity ID accessories
– Replacement interval monitoring
– Anti-interchange security for breathing hoses
For further information see: "Support of Infinity ID accessories", page 326.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Overview
49