Technical data
Accuracy
Inspiratory time Ti
Setting range
Increment value
Accuracy
Resulting ratio of inspiratory time to expiratory
time I:E
Maximum inspiratory time for assisted breaths
(not adjustable)
Pressure rise time Slope (T
Setting range
Increment value
Accuracy
Ratio of inspiratory time to expiratory time I:E
Setting range
Increment value
Accuracy
Inspiration termination criterion Insp term (as a
percentage of the peak inspiratory flow) for
pressure-supported breaths
Setting range
Increment value
Accuracy
Proportion of plateau time % Tplat to inspira-
tory time Ti for mandatory breaths in the VC -
CMV, VC - SIMV, and VC - SIMV / PS modes
Setting range
Increment value
Accuracy
Inflation time (Duration) including rising and
falling slopes in the One-step recruitment
maneuver
Setting range
Increment value
Accuracy
276
±10 % of the set value or ±1 /min (the greater value
applies)
0.2 to 10.0 s
0.1 s
±5 % of the set value or ±150 ms (the greater value
applies)
1:49 to 49:1
4.0 s for the "Adults" patient category
1.5 s for the "Pediatric patients" and "Neonates" patient
categories
time)
10...90
0.0 to 2.0 s
0.1 s
±25 % of the set value or ±250 ms (the greater value
applies)
4:1 to 1:10
0.1
The inspiratory time for the ventilator is calculated
based on the I:E and RR settings. The I:E accuracy
results from the accuracy of the inspiratory time, see
the information on Ti.
5 to 80 %
5 %
±5 % absolute error
0 to 60 %
10 %
±25 % absolute error
3 to 40 s
1 s
±10 % of the set value
Instructions for use
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n