Dräger Atlan A100 Instructions For Use Manual page 80

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Assembly and preparation
Connecting the vaporizers
WARNING
Risk due to improperly mounted vaporizers
Incorrectly mounted vaporizers can cause leakage. This can cause the fresh-gas
delivery to be too low or contaminate the ambient air. Patient and user can be
endangered.
► Make sure that the connected vaporizers are hanging vertically and are not tilted
by objects leaning against them.
► When using D-Vapor vaporizers, make sure that the power cable is not pinched.
► After mounting the vaporizers, perform a leakage test.
1. Set all the vaporizers upright and securely on the plug-in adapter.
2. Turn the locking levers (1) clockwise. The vaporizers are locked when the levers
are pointing to the left.
The illustration shows vaporizers of the type Dräger-Vapor 2000 with Auto
Exclusion.
WARNING
Risk due to incorrect anesthetic agent delivery
If the vaporizer is filled with the wrong anesthetic agent or if it is not filled sufficiently,
incorrect anesthetic gas concentrations or concentrations that are too low can occur
as a result.
► Compare the color code and labeling on the vaporizer used with the anesthetic
agent bottle.
► Observe the vaporizer filling level.
► Follow the instructions for use for the vaporizer.
3. Check the filling levels in the sight glasses (2). Fill the vaporizers if required.
4. Turn the control dial on each vaporizer to the 0 position. The key (3) engages.
Checking the metering interlock
The vaporizers have an interlock system which by means of a metering interlock
prevents the simultaneous opening of 2 vaporizers.
1. Turn the control dial on one of the vaporizers to a position other than 0.
80
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n