Backup manual mode with a mechanically controlled gas mixer with flow
tubes
1. Open the flap (1).
2. Activate the backup manual switch (2). Follow the instructions on the product
label (3) or the screen (4).
3. Ventilate the patient manually.
a. During mechanical ventilation or in Man/Spon ventilation mode:
b. When using the external fresh-gas outlet:
4. Set the fresh-gas flow.
5. Check the vaporizer setting.
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Ventilate manually with the breathing bag.
Ventilate manually with the breathing bag on the non-rebreathing system.