Dräger Atlan A100 Instructions For Use Manual page 240

Anesthesia workstation
Hide thumbs Also See for Atlan A100:
Table of Contents

Advertisement

Troubleshooting
12.8.1
Replacing the flow sensors
12.8.1.1
Removing the breathing system
To prevent accidental penetration of soda lime into the breathing system, make
sure that the reusable CO
Prerequisite: The breathing system cover has been removed.
1. To unlock, turn the 3 levers (1) counterclockwise by approximately 120°.
CAUTION
Risk of injury due to breathing system warmer
When the breathing system warmer is switched on, the bottom side of the breathing
system and the heating plate beneath it can become very hot.
► Allow the breathing system to cool off before removing.
2. Remove the breathing system (2) vertically upwards from the breathing system
mount.
240
absorber has been removed.
2
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n