Priority
Alarm
!
255
Inspiratory isoflurane low
!! 10
Inspiratory N
!
255
Inspiratory O
ment inaccurate
!! 90
Inspiratory pressure not
achieved
!!! 255
Inspiratory sevoflurane
high
!! 255
Inspiratory sevoflurane
high
!
255
Inspiratory sevoflurane
low
!! 50
Inspiratory tidal volume
high
!!! 255
Inspiratory xMAC high
!! 255
Inspiratory xMAC high
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Cause
The inspiratory anesthetic
gas concentration is below
the lower alarm limit.
The soda lime has dried out. Replace the soda lime.
O high
Inspiratory N
2
%.
measure-
The O
2
2
racy is reduced.
Fresh-gas flow is insuffi-
cient, breathing bag is
blocked or positioned incor-
rectly.
Leakage or disconnection.
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
Inspiratory anesthetic gas
concentration has exceeded
the upper alarm limit.
The inspiratory anesthetic
gas concentration is below
the lower alarm limit.
The soda lime has dried out. Replace the soda lime.
The ventilation settings are
not adequate for AutoFlow.
The inspiratory anesthetic
gas concentration has
exceeded 3 xMAC for more
than 30 seconds. The expira-
tory anesthetic gas concen-
tration has exceeded
2.5 xMAC for more than
30 seconds.
The inspiratory anesthetic
gas concentration has
exceeded 5 xMAC.
The inspiratory anesthetic
gas concentration has
exceeded 3 xMAC for more
than 180 seconds.
Remedy
Check vaporizer and fresh-
gas settings.
Refill the vaporizer.
Check the breathing system
for high leakage.
O exceeds 82
Check fresh-gas composi-
2
tion.
Press the O
the breathing system.
measurement accu-
When the case has been
completed, perform a sys-
tem test.
Check fresh-gas settings and
position of breathing bag.
Check the breathing circuit
for tight connections and
leakages.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Refill the vaporizer.
Check the breathing system
for high leakage.
Change the ventilation mode
or increase the tidal volume.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Check vaporizer and fresh-
gas settings.
Troubleshooting
+ button to flush
2
257