Manually triggering a breath
● Touch the Man. insp./Insp. hold button (3) briefly.
Manually extending a breath
1. Touch the Man. insp./Insp. hold button (3) and keep it pressed for the desired
time.
The device will trigger an extended breath or will extend an already triggered
automatic breath.
The breath is ended automatically:
– After a maximum of 40 seconds in the Adult patient category
– After a maximum of 30 seconds in the Ped patient category
– After a maximum of 5 seconds in the Neo patient category
8.5.2.2
Holding or extending the expiration
1. Touch the Exp. hold button (4) and keep it pressed for the desired time.
The device will extend the expiration and delay the next breath.
The expiration is ended automatically:
– After a maximum of 40 seconds in the Adult patient category
– After a maximum of 30 seconds in the Ped patient category
– After a maximum of 5 seconds in the Neo patient category
8.5.3
One-step recruitment
This maneuver applies a set pressure for a specific duration and thus enables, e.g.,
an extended breath with adjustable pressure.
1. Using the therapy controls, set the Pressure (5) and the Duration (6).
2. If necessary, touch the More button (7) and set a reminder with the therapy
control (8).
3. Touch the Start button (9) and confirm.
The rise in pressure from the PEEP level to the set pressure level and the decrease
in pressure at the end of the maneuver take place at 20 hPa/s.
|