8.7.3
Customizing the current view
Waveforms and parameter fields can be customized as follows:
1. Touch the waveform or parameter field.
The Quick setup dialog opens.
WARNING
Risk due to inadequate monitoring
National and medical regulations may require certain parameters to be displayed.
► Always consider the relevant regulations when configuring the screen layout.
2. For Content (1), select the desired content.
For a list of the possible screen content, see page 225.
3. For Scale (2), select the desired setting.
Unsaved changes to the current view are kept until the next device restart.
8.7.3.1
Restoring the current view
The changes to the current view can be canceled.
1. Open the Views dialog.
2. Touch the Restore current view button.
8.7.4
Using loops
Prerequisite: The device has the "advanced ventilation monitoring" or "loops and
trends" software option.
The following loops are available:
– Pressure-Volume loop
– Flow-Volume loop
|
Instructions for use
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
Quick setup
2
1
Operation
171