Operation
8.3.1.1
Manual ventilation
● Set the APL valve to the desired maximum airway pressure.
The patient can be ventilated with the breathing bag. The pressure is limited to
the set value.
● To release pressure from the breathing system in Manual / Spontaneous mode,
lift the valve.
8.3.1.2
Spontaneous breathing
● Turn the APL valve counterclockwise as far as it will go.
The dots are aligned vertically over one another. The valve lifts.
The pressure limitation is canceled and the valve is open for free spontaneous
breathing.
8.3.2
Using the O
The O
breathing bag with oxygen. The vaporizers are bypassed for this.
● Press the O
O
148
10
20
flush
2
flush is used for flushing and quickly filling the breathing system and
2
+ key.
2
continues to flow for as long as the key is pressed.
2
The gas concentration can change abruptly when the O
Instructions for use
30
|
Atlan A100 (XL), A300 (XL), A350 (XL) SW 2.1n
flush is used.